Tehran University of Medical Sciences

Science Communicator Platform

Stay connected! Follow us on X network (Twitter):
Share this content! On (X network) By
Reduced Imaging Rate in Liver Elastometery Using Shear Wave Interference Patterns Publisher Pubmed



Soozande M1 ; Arabalibeik H1, 2 ; Alavian SM3
Authors
Show Affiliations
Authors Affiliations
  1. 1. Department of Medical Physics and Biomedical Engineering, Tehran University of Medical Sciences, Tehran, Iran
  2. 2. Research Center of Biomedical Technology and Robotics (RCBTR), Tehran University of Medical Sciences, IK Hospital Complex, Keshavarz Blvd., Tehran, Iran
  3. 3. Center for Gastroenterology and Liver Disease, Baqiyatallah University of Medical Sciences, Tehran, Iran

Source: Studies in Health Technology and Informatics Published:2016


Abstract

Inducing interference patterns of shear wave is one of the proposed methods for reducing the frame rate in measuring wave speed during tissue elastography. Previously, the Nyquist rate must be met in order to provide an appropriate image for extracting the patterns with a reasonable accuracy. In this article we propose a technique based on image registration, and apply it to ultrasound images acquired before and after inducing the shear waves to estimate the amplitude of displacement. The displacement of the tissue is then used to form the interference pattern of shear waves. The method does not induce any restrictions on the time interval between images, so the tissue elasticity can be calculated independent of the imaging rate. The average error in measuring the elasticity of the simulated phantom is 13.7%. © 2016 The authors and IOS Press. All rights reserved.