Tehran University of Medical Sciences

Science Communicator Platform

Stay connected! Follow us on X network (Twitter):
Share this content! On (X network) By
Pneumothorax Caused by Anesthesia Circuit Misconnection



Shahriari A1 ; Khooshideh M2 ; Sheikh M3
Authors
Show Affiliations
Authors Affiliations
  1. 1. Departement of Anesthesiology, Roozbeh Hospital, Tehran University of Medical Sciences, Tehran, Iran
  2. 2. Departement of Obstetrics and Gynecology, Arash Hospital, Tehran University of Medical Sciences, Tehran, Iran
  3. 3. Maternal, Fetal and Neonatal Research Center, Tehran University of Medical Sciences, Vali-asr Hospital, Imam Khomeini Hospital Complexes, Keshavarz Blvd., Tehran, 1419733141, Iran

Source: Anaesthesia# Pain and Intensive Care Published:2016

Abstract

Preoperative checking of the anesthetic equipment which is the primary responsibility of the nurse anesthetist is crucial for patient safety. There are many misconnection possibilities with circle system that can contribute to as highly as 35% of adverse anesthetic outcomes. A 32-year old man was admitted to the operating room (OR) for appendectomy. After the induction of anesthesia and beginning mask ventilation the anesthesiologist noticed that ventilation was not effective, therefore performed endotracheal intubation. The tracheal tube was connected to the Y piece and the mechanical ventilation was started by anesthesia machine. The lungs were not expanding effectively, therefore manual ventilation was began by an ambu bag. Beginning the manual ventilation the anesthesiologist realized an increased resistance against ventilation and an elevated airways pressure. The Sp02 fell to 79% and tachycardia was evident. Physical examinations were suggestive of pneumothorax that occurred within the first 4-5 minutes of starting machine ventilation. Rechecking the anesthesia machine revealed an error in the connection of the anesthesia circuit. We briefly review the current literature and give suggestions to eliminate these mismanagements in the operating room which can be life threatening.