Comparison of the Accuracy of Anterior Segment Optical Coherence Tomography and Ultrasound Biomicroscopy in Localizing Rectus Muscle Insertions Publisher Pubmed



Mirmohammadsadeghi A1 ; Hamzeh N1 ; Ghassemi F1 ; Akbari MR1
Authors
Show Affiliations
Authors Affiliations
  1. 1. Eye Research Center, Farabi Eye Hospital, Tehran University of Medical Sciences, Tehran, Iran

Source: Journal of Binocular Vision and Ocular Motility Published:2022


Abstract

Purpose: To compare the accuracy of anterior segment optical coherence tomography (AS-OCT) and ultrasound biomicroscopy (UBM) in localizing rectus muscle insertions. Methods: The study was performed on 27 patients (39 rectus muscles) who required primary or secondary surgery. Using caliper function in the AS-OCT and UBM software, the distance from the insertion site to the anterior chamber angle was measured. The actual muscle insertion distance from limbus was considered as the measured distance plus 1 mm. The measurements by UBM and AS-OCT were compared with intraoperative measurements and with each other. Results: AS-OCT and UBM were performed on 13 medial rectus, 24 lateral rectus, and 2 superior rectus muscles. Ninety two percent of UBM measurements (36 muscles) were within 1 mm, one was within 1–1.5 mm, and 2 were within 1.5–2 mm of surgery measurements. Eighty five percent of AS-OCT measurements (33 muscles) were within 1 mm, 5 were within 1–1.5 mm, and one was within 1.5–2 mm of surgery measurements. In all cases, the mean absolute error of the UBM (0.54 ± 0.44) and AS-OCT (0.51 ± 0.36) showed no significant difference (p = .76). Conclusion: AS-OCT and UBM can be used interchangeably to localize rectus muscle insertions and showed good agreement with intraoperative measurements. © 2022 American Orthoptic Journal Inc.