Tehran University of Medical Sciences

Science Communicator Platform

Stay connected! Follow us on X network (Twitter):
Share this content! On (X network) By
Entrapped Arterial Sheath During Transfemoral Angiography Publisher



Salimi J1 ; Majd A1 ; Hosseini S2 ; Shariatzadeh S1 ; Behzadi M1
Authors
Show Affiliations
Authors Affiliations
  1. 1. Sina Hospital, Department of Surgery, Tehran University of Medical Sciences, Tehran, Iran
  2. 2. Shahid Rajaie Cardiovascular, Medical and Research Center, Tehran, Iran

Source: Journal of Pediatric Surgery Case Reports Published:2021


Abstract

Arterial sheath entrapment is a potential complication during Trans Femoral Angiography. Here we are presenting a nine-month-old boy, a candidate for surgical repair of Tetralogy of Fallot, with complicated sheath entrapment. As the problem was not resolved initially by pharmacological approaches, later a transverse incision on the artery was made at the mid-length of the sheath. Subsequently, the device was splitted into two halves and both were successfully removed through the incision. This method facilitated the sheath withdrawal and diminished the trauma applied to the artery during surgery. © 2021