Tehran University of Medical Sciences

Science Communicator Platform

Stay connected! Follow us on X network (Twitter):
Share this content! On (X network) By
Evaluation of Annual Staff Doses and Radiation Shielding Efficiencies of Thyroid Shield and Lead Apron During Preparation and Administration of 131 I, 81 Kr, and 99M Tc-Labeled Radiopharmaceuticals Publisher



Khalili N1 ; Zakariaee SS2 ; Gharebaghi EJ1 ; Salehi Y3 ; Changizi V1
Authors
Show Affiliations
Authors Affiliations
  1. 1. Department of Radiology and Radiotherapy, Allied Medical Sciences School, Tehran University of Medical Sciences, Tehran, Iran
  2. 2. Department of Medical Physics, Faculty of Paramedical Sciences, Ilam University of Medical Sciences, Ilam, Iran
  3. 3. Department of Nuclear Medicine, Vali-Asr Hospital, Tehran University of Medical Sciences, Tehran, Iran

Source: Journal of Medical Signals and Sensors Published:2022


Abstract

Nuclear medicine technicians would receive unavoidable exposures during the preparation and administration of radiopharmaceuticals. Based on the staff dose monitoring, the dose reduction efficiencies of the radiation protection shields and the need to implement additional strategies to reduce the staff doses could be evaluated. In this study, medical staff doses during the preparation and administration of Tc-99 m, I-131, and Kr-81 radiopharmaceuticals were evaluated. The dose reduction efficiencies of the lead apron and thyroid shield were also investigated. GR-207 thermoluminescent dosimeter (TLD) chips were used for quantifying the medical staff doses. The occupational dose magnitudes were determined in five organs at risk including eye lens, thyroid, fingers, chest, and gonads. TLDs were located under and over the protective shields for evaluating the dose reduction efficiencies of the lead apron and thyroid shield. The occupational doses were normalized to the activities used in the working shifts. During preparation and injection of Tc-99 m radiopharmaceutical, the average annual doses were higher in the chest (4.49 mGy) and eye lenses (4 mGy). For I-131 radiopharmaceutical, the average annual doses of the point-finger (15.8 mGy) and eye lenses (1.23 mGy) were significantly higher than other organs. During the preparation and administration of Kr-81, the average annual doses of the point-finger (0.65 mGy) and chest (0.44 mGy) were higher. The significant dose reductions were achieved using the lead apron and thyroid shield. The radiation protection shields and minimum contact with the radioactive sources, including patients, are recommended to reduce the staff doses. © 2021 Journal of Medical Signals & Sensors Published by Wolters Kluwer-Medknow.