Relationship of the Optic Nerve Sheath Diameter and Repeated Invasive Intracranial Pressure Measures in Traumatic Brain Injury Patients; a Diagnostic Accuracy Study Publisher



Mehrpour S1 ; Najafi A2 ; Ahmadi A2 ; Zarei T3 ; Pleqi V4 ; Basiri K5, 6 ; Komlakh K7 ; Abdollahi H2 ; Emami KH2
Authors
Show Affiliations
Authors Affiliations
  1. 1. Department of Anesthesiology and Critical Care, Qom University of Medical Sciences, Qom, Iran
  2. 2. Department of Anesthesiology and Critical Care, Sina Hospital, Tehran University of Medical Sciences, Tehran, Iran
  3. 3. Department of Anesthesiology and Critical Care, Hormozgan University of Medical Sciences, Bandar Abbas, Iran
  4. 4. Department of Clinical Research, Mayo Clinic, Jacksonville, FL, United States
  5. 5. Prehospital and Hospital Emergency Research Center, Tehran University of Medical Sciences, Tehran, Iran
  6. 6. Department of Emergency Medicine, Sina Hospital, Tehran University of Medical Sciences, Tehran, Iran
  7. 7. Department of Neurosurgery, Imam Hossein Hospital, Shahid Beheshti University of Medical Sciences, Tehran, Iran

Source: Frontiers in Emergency Medicine Published:2022


Abstract

Objective: The purpose of this study was to quantitatively evaluate if the use of the optic nerve sheath diameter (ONSD) can be a suitable noninvasive surrogate approach for repeated invasive intracranial pressure (ICP) measures. Methods: The study used a sample of 22 adult patients with traumatic brain injury (TBI) from an in intensive care unit (ICU). ICP levels were measured using the gold standard and recorded in cmH20. ONSD was measured using ultrasonography with 5.6-5.7 MHz linear probe and recorded in millimeters. The data analysis was done using STATA software version 15. Results: The results showed a strong positive correlation between ICP and ONSD (r = 0.743, p = 0.001). The accuracy of the sonographic ONSD declined over time, starting from a high of 90.9% at the baseline and declining to a low of merely 20.0% after 48 hours. Conclusion: These findings indicate that the ONSD approach could be very useful alternative and noninvasive method for monitoring ICP. © 2022 Tehran University of Medical Sciences.