Applicability of Cone Beam Computed Tomography Gray Values for Estimation of Primary Stability of Dental Implants; [Aplicabilidade Dos Valores De Cinza Da Tomografia Computadorizada De Feixe Conico Na Estimativa Da Estabilidade Primaria De Implantes Dentarios] Publisher



Sheikhi M1 ; Karami M2 ; Abbasi S3 ; Moaddabi A4 ; Soltani P5
Authors
Show Affiliations
Authors Affiliations
  1. 1. Torabinejad Dental Research Center, Department of Oral and Maxillofacial Radiology, School of Dentistry, Isfahan University of Medical Sciences, Isfahan, Iran
  2. 2. Department of Radiology, School of Medicine, Isfahan University of Medical Sciences, Isfahan, Iran
  3. 3. Department of Mathematics, Isfahan (khorasgan) Branch, Islamic Azad University, Isfahan, Iran
  4. 4. Department of Oral and Maxillofacial Surgery, School of Dentistry, Mazandaran University of Medical Sciences, Sari, Iran
  5. 5. Department of Oral and Maxillofacial Radiology, Dental Implants Research Center, Dental Research Institute, Isfahan University of Medical Sciences, Isfahan, Iran

Source: Brazilian Dental Science Published:2021


Abstract

Objective: To evaluate the relationship between bone density measured by computed tomography (CT) and cone beam computed tomography (CBCT) (Sirona’s Galileos scanner) with primary stability of dental implants. Material and methods: 20 fresh bovine femoral heads were prepared by removal of soft tissue, sectioning of the bone, and placement of markers for location and angulation of implants. Bone density of peri-implant areas was determined preoperatively by CT and CBCT scanning of the prepared bone samples represented by Hounsfield units (HUs) and gray values (GVs), respectively. Then, 60 implants in three sizes (diameter = 4 mm, length = 8, 10, and 12 mm) were inserted into the bones and maximum insertion torque (IT) was recorded. Osstell device was also used for determining the implant stability quotient (ISQ) for each implant. Statistical analysis was performed on the data (α = 0.05). Results: Mean ± SD values of GV, HU, and ISQ were 1592.75 ± 231.82, 675.26 ± 115.38 and 61.90 ± 10.14, respectively. Moreover, the most frequent IT limit was 30-35 Ncm (41.4%). Significant relationships were observed between HU and IT, GV and IT, HU and ISQ, GV and ISQ, and IT and ISQ in all implant sizes. Moreover, GV and HU also significantly correlated to each other. Conclusion: Bone density values in CBCT and CT scans are positively associated to primary stability of dental implants. Therefore, GVs obtained from Galileos CBCT scanner can be used for preoperative selection of edentulous sites which allow for better implant stability or locations which require further procedures for enhancing the success rate of dental implants. © 2021, Universidade Estadual Paulista, Institute of Science and Technology of Sao Jose dos Campos. All rights reserved.
Other Related Docs
16. Designing an Optimized Novel Femoral Stem, Journal of Medical Signals and Sensors (2017)
18. Accuracy of Tooth-Implant Impressions: Comparison of Five Different Techniques, Clinical and Experimental Dental Research (2023)
29. Buckling Analysis of a Functionally Graded Implant Model for Treatment of Bone Fractures: A Numerical Study, ASME International Mechanical Engineering Congress and Exposition, Proceedings (IMECE) (2017)
39. Bone Loss-Related Factors in Tissue and Bone Level Dental Implants: A Systematic Review of Clinical Trials, Journal of the Korean Association of Oral and Maxillofacial Surgeons (2021)