Evaluation of Dosimetric Characteristics of Diodes and Ionization Chambers in Small Megavoltage Photon Field Dosimetry Publisher



Keivan H1, 2 ; Shahbazigahrouei D1 ; Shanei A1
Authors
Show Affiliations
Authors Affiliations
  1. 1. Department of Medical Physics, School of Medicine, Isfahan University of Medical Sciences, Isfahan, Iran
  2. 2. School of Paramedicine, Shahroud University of Medical Sciences, Shahroud, Iran

Source: International Journal of Radiation Research Published:2018


Abstract

Background: Modern radiaon therapy such as intensity modulated radiaon therapy (IMRT) and volumetric modulated arc therapy (VMAT) has resulted in using small therapeuc photon beams. The scope of this work is to invesgate the variaon in efficiency of different ionizaon chambers and semiconductor diodes in small mul-leaf collimator (MLC) defined fields of ARTISTE linear accelerator. Materials and Methods: Dose distribuons measurements was done in field sizes ranging from 0.5×0.5 to 10×10 cm2 combining with Monte Carlo (MC) simulaon. The treatment head of linac and the detectors were simulated by means of BEAMnrc/DOSXYZnrc of EGSnrc MC. The parameters such as output rao (OR), penumbral width, dosimetric field size and the percentage surface dose in small radiaon fields was evaluated using ionizaon chambers and semiconductor dosimeters. Results: ORs and beam profiles resulng from various measurements showed significant difference between ionizaon chambers and diodes in small fields. Derivaon of less than 2% was observed between EDGE and Diode P. ORs vary by more than 30% for 1×1 cm2 field size but, in larger field sizes differences was less than 1%. Penumbra underesmaon was seen in Semiflex relave to pinpoint ionizaon chamber. No difference was seen in the pa?ern of surface dose among the applied detectors. Conclusion: Dosimetric characteriscs of different detectors showed significant differences in small photon beams. Profiles and ORs analysis with different dosimeters showed that not only water equivalency of detectors, but also dosimeter acve volume is important factors for determinaon of dosimetric behavior in small photon beams. © 2018 Novin Medical Radiation Institute.All right reserved.
Other Related Docs
17. Sensitivity Analysis of a 6 Mev Photon Beam Monte Carlo Model, Journal of Medical Signals and Sensors (2023)
35. Monte Carlo Study of Fetal Dosimetry Parameters for 6 Mv Photon Beam, Journal of Medical Signals and Sensors (2013)
37. Cisplatin-Induced Renal Toxicity: A Short Review, Life Science Journal (2014)