Impact of Magnetic Fields on Calculated Aapm Tg-43 Parameters for 192Ir and 60Co Hdr Brachytherapy Sources: A Monte Carlo Study Publisher Pubmed



Ranjbar S1 ; Gholami S2, 3 ; Longo F4 ; Sharp R5 ; Meigooni AS6
Authors
Show Affiliations
Authors Affiliations
  1. 1. Department of Medical Physics and Engineering, School of Medicine, Shiraz University of Medical Sciences, Shiraz, Iran
  2. 2. Radiation Oncology Research Center, Cancer Institute, Tehran University of Medical Sciences, Tehran, Iran
  3. 3. Department of Medical Physics and Biomedical Engineering, Tehran University of Medical Sciences, Tehran, Iran
  4. 4. Department of Physics, University of Trieste and INFN Trieste, Italy
  5. 5. Department of Health Physics and Diagnostic Sciences, University of Nevada, Las Vegas, NV, United States
  6. 6. Comprehensive Cancer Centers of Nevada, Las Vegas, NV, United States

Source: Applied Radiation and Isotopes Published:2020


Abstract

Purpose: The aim of this work is to investigate the influence of an external magnetic field (MF) on The American Association of Physicists in Medicine (AAPM) No. 43 Report (TG-43) parameters for 192Ir and 60Co high dose rate (HDR) brachytherapy sources using Monte Carlo (MC) simulation methods. Materials and methods: We used the Geant4 toolkit (version 10.1. p01) to simulate the geometry of 192Ir and 60Co brachytherapy sources. AAPM TG-43 parameters (the radial dose function, g(r), and the anisotropy function, F (r, θ)) of both 192Ir and 60Co sources were calculated in the presence of a magnetic field with strengths of 1.5T, 3T, and 7T in the X, Y, and Z directions in a voxelized water phantom. Results: For the 192Ir source, the calculated values g(r) and F (r, θ) remained nearly unaffected by the magnetic field for all investigated strengths. For the 60Co source, the differences for the g(r) and F (r,θ) under the 1.5T, 3T, and 7T magnetic field strengths along the direction parallel with the MF were found to be an increase of up to 5%, 15%, and 33%, respectively. However, for the directions perpendicular with the magnetic field, there was a decrease of up to 3%, 6% and 15% under 1.5T, 3T and 7T strengths, respectively. Conclusion: Our results highlight the necessity of a Monte Carlo-based treatment planning system (TPS) if cobalt HDR treatments are performed under a magnetic field, especially for strengths greater than 1.5T. © 2020 Elsevier Ltd