Pragmatic Approaches to Reducing Radiation Dose in Brain Computed Tomography Scan Using Scan Parameter Modification Publisher



Choopani MR1 ; Abedi I2 ; Dalvand F3
Authors
Show Affiliations
Authors Affiliations
  1. 1. Departments of Radiology, Isfahan University of Medical Sciences, Isfahan, Iran
  2. 2. Department of Medical Physics, Isfahan University of Medical Sciences, Isfahan, Iran
  3. 3. Department of Medical Radiation Engineering, Shahid Beheshti University, Tehran, Iran

Source: Journal of Medical Signals and Sensors Published:2022


Abstract

Background: High radiation dose of patients has become a concern in the computed tomography (CT) examinations. The aim of this study is to guide the radiology technician in modifying or optimizing the underlying parameters of the CT scan to reduce the patient radiation dose and produce an acceptable image quality for diagnosis. Methods: The body mass measurement device phantom was repeatedly scanned by changing the scan parameters. To analyze the image quality, software-based and observer-based evaluations were employed. To study the effect of scan parameters such as slice thickness and reconstruction filter on image quality and radiation dose, the structural equation modeling was used. Results: By changing the reconstruction filter from standard to soft and slice thickness from 2.5 mm to 5 mm, low-contrast resolution did not change significantly. In addition, by increasing the slice thickness and changing the reconstruction filter, the spatial resolution at different radiation conditions did not significantly differ from the standard irradiation conditions (P > 0.05). Conclusion: In this study, it was shown that in the brain CT scan imaging, the radiation dose was reduced by 30%-50% by increasing the slice thickness or changing the reconstruction filter. It is necessary to adjust the CT scan protocols according to clinical requirements or the special conditions of some patients while maintaining acceptable image quality. © 2022 Journal of Medical Signals & Sensors.
Experts (# of related papers)
Other Related Docs
9. Optimization of Ct Slice Thickness in 3D-Crt and Imrt Planning of Low Grade Glioma, International Journal of Radiation Research (2021)
10. Patient Dose Resulting From Ct Examinations in Yazd, Iran, Iranian Journal of Radiation Research (2006)
25. Monte Carlo Study of Fetal Dosimetry Parameters for 6 Mv Photon Beam, Journal of Medical Signals and Sensors (2013)
36. Children, Ct Scan and Radiation, International Journal of Preventive Medicine (2010)
39. Dose Response Evaluation of a Low Density Anoxic Polymer Gel Dosimeter Using Mri, International Journal of Radiation Research (2015)
48. Evaluation of Radiographers' and Ct Technologists' Knowledge Regarding Ct Exposure Parameters, Polish Journal of Medical Physics and Engineering (2019)