Evaluation of Exit Skin Dose for Intra-Cavitary Brachytherapy Treatments by the Bebig 60Co Machine Using Thermoluminescent Dosimeters Publisher



Marvi M1 ; Gholami S2 ; Barough MS1 ; Hosseini M1 ; Nabavi M2 ; Jaberi R2 ; Mohammadkarim A3
Authors
Show Affiliations
Authors Affiliations
  1. 1. Department of Medical Radiation Engineering, Central Tehran Branch, Islamic Azad University, Tehran, Iran
  2. 2. Radiation Oncology Department, Cancer Institute, Tehran University of Medical Sciences, Tehran, Iran
  3. 3. Department of Medical Physics, Semnan University of Medical Sciences, Semnan, Iran

Source: Journal of Radiotherapy in Practice Published:2021


Abstract

Purpose: This study aims to evaluate the application of the exit skin dose (ESD) in verifying the accuracy of intra-cavitary brachytherapy treatments performed by the BEBIG 60Co machine using thermoluminescent dosimeters (TLDs). Materials and methods: Eleven patients who were treated for gynaecological (GYN) malignancy by high-dose-rate (HDR) brachytherapy machine have been considered in this study. A combination of tandem, cylinder and interstitial needles was applied for eight patients while tandem ovoid (TO) applicators were used for the rest (three patients). In order to measure ESD, thermoluminescent dosimetry was performed for each patient. TLDs were placed precisely on the patient's skin along her symphysis pubis bone (anterior) and left (L)/right (R) sides of her pelvic. Positioning of the dosimeter was accurately determined using fiducial markers in computed tomography (CT) scan imaging, prior to the treatment. Finally, a comparison was made between the calculated dose from the treatment planning system (TPS) and the dose measured by TLDs. Results: About 90% of all cases showed a good agreement (while considering TLD uncertainty ∼5·5%) between TPS dose calculations and TLD measurements. The measured mean values of ESD received to anterior, left and right positions were 56·72, 12·18 and 12·82 cGy, respectively. For three patients, differences up to 11·9% were detected. Conclusion: To conclude, ESD measurement method can be a suitable practical approach for verifying the accuracy of GYN HDR treatment delivery. ©