Precise Ebt3 Gafchromic Film Dosimetry for Grid Therapy Publisher



Gholami S1, 2 ; Nedaie HA1, 2 ; Meigooni AS3
Authors
Show Affiliations
Authors Affiliations
  1. 1. Radiotherapy Oncology Research Centre, Cancer Institute, Tehran University of Medical Sciences, Tehran, Iran
  2. 2. Department of Medical Physics and Biomedical Engineering, Tehran University of Medical Sciences, Tehran, Iran
  3. 3. Comprehensive Cancer Centers of Nevada, Las Vegas, NV, United States

Source: Radiation Measurements Published:2019


Abstract

Objectives: Grid therapy is a radiation therapy technique for the treatment of bulky tumours. The high dose gradients and non-uniformity of dose distributions within the target lead to a challenge in the dosimetry of the Grid radiation fields. The aim of this study is to perform a precise EBT3 Gafchromic film dosimetry for Grid therapy fields using a commercially available flatbed colour scanner. Methods: In this project, samples of the EBT3 Gafchromic films are exposed to Grid radiation fields. The irradiated EBT3 films were read using a flatbed Microtek scanner. The responses of these films (i.e. films from the same batch) as functions of the absorbed dose values are calibrated by irradiation under a fixed standard technique (i.e. 10 × 10 cm2 filed, 100 cm SSD, and depth of the maximum dose). These films are also read with the same scanner using the red, green, and blue channels. Four different approaches were used to evaluate film dosimetry for the Grid therapy applications: 1) single channel film dosimetry method (SCM), 2) dual channel film dosimetry method (DCM), 3) linearized dose-response curve method (LRCM), and 4) triple channel film dosimetry method (TCM). A dose of 20 Gy was delivered to the point along the central axis of the grid hole at the depth of maximum dose (dmax) for a 20 × 20 cm2 Grid field size. Beam profiles and percentage depth dose distributions of the Grid radiation have been measured in water-equivalent phantom material, using EBT3 films. The accuracy of the relative and the absolute dosimetry of the films were examined by comparison of the TLD measured data with the Monte Carlo simulated values. Results: The results of these investigations show that for a gamma index criterion of 5%/3 mm, the agreements between the MC calculations dose profiles and the SCM, DCM, LRCM, and TCM film dosimetry approaches the passing rates of 91%, 92%, 95%, and 95%, respectively. A much closer agreement was observed for using a linearized dose-response curve and triple-channel methods. Conclusions: Selection of an appropriate methodology in Gafchromic film dosimetry may lead to an accurate dose-response in a high dose gradient radiation field such as Grid therapy. © 2018 Elsevier Ltd
Other Related Docs
21. Management of Bulky High-Grade Pleomorphic Sarcoma Using Grid Therapy Technique, International Journal of Radiation Research (2021)
27. Different Dosimeters/Detectors Used in Small-Field Dosimetry: Pros and Cons, Journal of Medical Signals and Sensors (2018)