Monte Carlo Optimization of Crystal Configuration for Pixelated Molecular Spect Scanners Publisher



Mahani H1, 2 ; Raisali G1 ; Kamaliasl A3 ; Ay MR2, 4
Authors
Show Affiliations
Authors Affiliations
  1. 1. Radiation Application Research School, Nuclear Science and Technology Research Institute, Tehran, Iran
  2. 2. Research Center for Molecular and Cellular Imaging, Tehran University of Medical Science, Tehran, Iran
  3. 3. Radiation Medicine Department, Shahid Beheshti University, Tehran, Iran
  4. 4. Department of Medical Physics and Biomedical Engineering, Tehran University of Medical Science, Tehran, Iran

Source: Nuclear Instruments and Methods in Physics Research# Section A: Accelerators# Spectrometers# Detectors and Associated Equipment Published:2017


Abstract

Resolution-sensitivity-PDA tradeoff is the most challenging problem in design and optimization of pixelated preclinical SPECT scanners. In this work, we addressed such a challenge from a crystal point-of-view by looking for an optimal pixelated scintillator using GATE Monte Carlo simulation. Various crystal configurations have been investigated and the influence of different pixel sizes, pixel gaps, and three scintillators on tomographic resolution, sensitivity, and PDA of the camera were evaluated. The crystal configuration was then optimized using two objective functions: the weighted-sum and the figure-of-merit methods. The CsI(Na) reveals the highest sensitivity of the order of 43.47 cps/MBq in comparison to the NaI(Tl) and the YAP(Ce), for a 1.5×1.5 mm2 pixel size and 0.1 mm gap. The results show that the spatial resolution, in terms of FWHM, improves from 3.38 to 2.21 mm while the sensitivity simultaneously deteriorates from 42.39 cps/MBq to 27.81 cps/MBq when pixel size varies from 2×2 mm2 to 0.5×0.5 mm2 for a 0.2 mm gap, respectively. The PDA worsens from 0.91 to 0.42 when pixel size decreases from 0.5×0.5 mm2 to 1×1 mm2 for a 0.2 mm gap at 15° incident-angle. The two objective functions agree that the 1.5×1.5 mm2 pixel size and 0.1 mm Epoxy gap CsI(Na) configuration provides the best compromise for small-animal imaging, using the HiReSPECT scanner. Our study highlights that crystal configuration can significantly affect the performance of the camera, and thereby Monte Carlo optimization of pixelated detectors is mandatory in order to achieve an optimal quality tomogram. © 2016 Elsevier B.V.