Global and Spatial Dosimetric Characteristics of N-Vinylpyrrolidone-Based Polymer Gel Dosimeters As a Function of Medium-Term Post-Preparation and Post-Irradiation Time Publisher



Farahani S1 ; Moslehshirazi MA1, 2 ; Riyahi Alam N3, 4 ; Rabi Mahdavi S5 ; Raeisi F1
Authors
Show Affiliations
Authors Affiliations
  1. 1. Ionizing and Non-ionizing Radiation Protection Research Center (INIRPRC), School of Paramedical Sciences, Shiraz University of Medical Sciences, Shiraz, Iran
  2. 2. Physics Unit, Department of Radio-oncology, School of Medicine, Shiraz University of Medical Sciences, Shiraz, Iran
  3. 3. Department of Medical Physics and Biomedical Engineering, Tehran University of Medical Sciences, Tehran, Iran
  4. 4. PERFORM Center, Concordia University, Montreal, QC, Canada
  5. 5. Department of Medical Physics, School of Medicine, Iran University of Medical Sciences, Tehran, Iran

Source: Radiation Physics and Chemistry Published:2022


Abstract

It is useful in practice to know the best time intervals between gel preparation, irradiation and imaging for optimal dosimetric response. Knowing how long after preparation and/or irradiation a gel dosimeter can still produce reliable results is also practically helpful. This study aimed to answer these two important but little-investigated questions for normoxic N-vinylpyrrolidone-based polymer gel dosimeters (VIPET) during a nine-day time period. We evaluated the global dose response, reproducibility and high-dose-gradient spatial integrity of this gel dosimetry system for up to 6 Gy dose and assessed the medium-term post-preparation and post-irradiation stability of these key properties with relatively short time intervals of 3 days. After synthesis, several batches of VIPET vials were irradiated with 6 MV photon beams and imaged on a 3 T magnetic resonance imaging (MRI) scanner in different time and dose schemes and were subsequently analysed. Statistical analysis of the experimental data showed acceptable relative dosimetric characteristics in terms of dose sensitivity, offset (R0) values, and linearity of synthesized gel over the course of a nine-day post-preparation and post-irradiation time. The results showed reasonably good reproducibility of dose sensitivity (standard error of measurement, SEM: 0.009) and R0 value (SEM: 0.023). Further, the measurements indicated satisfactory spatial stability of relative dose distribution in high-dose-gradient regions throughout the nine-day studied period (variations in beam penumbra widths and off-axis positions being typically less than 1 mm). These findings provide further evidence to the suitability of the VIPET gel for practical three-dimensional dosimetry in clinical settings such as modern radiotherapy. © 2022 Elsevier Ltd
Other Related Docs
9. Different Dosimeters/Detectors Used in Small-Field Dosimetry: Pros and Cons, Journal of Medical Signals and Sensors (2018)
10. Evaluation of Mri-Based Magic Polymer Gel Dosimeter in Small Photon Fields, International Journal of Radiation Research (2021)
12. Presage® As a Solid 3-D Radiation Dosimeter: A Review Article, Radiation Physics and Chemistry (2017)
16. Evaluation of Mri-Based Magic Polymer Gel Dosimeter in Small Photon Fields, International Journal of Radiation Research (2016)
18. Validation of a Prototype Optical Computed Tomography System, Journal of Medical Signals and Sensors (2015)