Development and Characterization of a Compact Hand-Held Gamma Probe System, Surgeoguide, Based on Nema Nu3-2004 Standards Publisher



Kaviani S1 ; Zeraatkar N1 ; Sajedi S1 ; Gorjizadeh N1, 2 ; Farahani MH1 ; Ghafarian P3, 4 ; Fakhri GE5 ; Sabet H5 ; Ay MR1, 2
Authors
Show Affiliations
Authors Affiliations
  1. 1. Research Center for Molecular and Cellular Imaging, Tehran University of Medical Sciences, Keshavarz Blvd, Tehran, Iran
  2. 2. Department of Medical Physics and Biomedical Engineering, Tehran University of Medical Sciences, Pour Sina, Tehran, Iran
  3. 3. Chronic Respiratory Diseases Research Center, National Research Institute of Tuberculosis and Lung Diseases (NRITLD), Shahid Beheshti University of Medical Sciences, Niyavaran, Tehran, Iran
  4. 4. PET/CT and Cyclotron Center, Masih Daneshvari Hospital, Shahid Beheshti University of Medical Sciences, Niyavaran, Tehran, Iran
  5. 5. Massachusetts General Hospital, Harvard Medical School, 55 Fruit, Boston, United States

Source: Journal of Instrumentation Published:2016


Abstract

Using an intra-operative gamma probe after periareolar or peritumoral injection of a radiotracer during surgery helps the surgeon to identify the sentinel, or first, nodal site of regional metastasis in clinically node-negative patients. The pathological analysis of this node can have an important influence on the treatment staging in various cancers. This paper reports the design and performance evaluation of a gamma probe recently developed in our department. The detector unit of this system consists of an 8 mm diameter and 10 mm thickness monolithic CsI(Tl) scintillator optically, coupled to a Silicon Photomultiplier (SiPM) with an active area of 6×6 mm2, and a single-hole collimator. The unit is shielded using tungsten. The system can operate in three different modes for Tc-99m, I-131, or F-18 isotopes. The following measurements were carried out to evaluate the performance of the probe: sensitivity in air and scatter medium, spatial resolution in scatter medium, angular resolution in scatter medium, and side and back shielding effectiveness. All experiments have been performed based on the NEMA NU3-2004 standard set up. The measured system sensitivities in air and scatter medium (water) are 1700 cps/MBq and 1770 cps/MBq, respectively, both measured at 3 cm from the collimator. The spatial resolution in the scatter medium is about 45 mm at 3 cm distance from the collimator. Also, the angular resolution of the probe is 74 FWHM. Finally, a shielding effectiveness of 99.5% is measured. The results show that the probe can potentially be used for sentinel lymph node localization during the surgery. © 2016 IOP Publishing Ltd and Sissa Medialab srl.